MultiPath Vectorial VNA Calibration
MultiPath Vectorial VNA Calibration allows to correct the VNA receivers measurement in multiple DUT's planes. Multiple VNA calibration is available when the VNA and I/O switch instruments are activated in the Setup Editor. Depending on the measurement type and the level of accuracy needed for the measurements, different calibration are possible.
VNA Cal Type | Setup Editor Requirements | Calibration Step | Available Measurements Output data |
---|---|---|---|
Full 2-Ports Calibration | 1) Vector Network Analyzer and I/O Switchenabled 2) Number of VNA receivers set to 4 |
12 error terms matrix extraction from TRL, SOLT, SOLR, … to correct Ratioed Measurements | S-Parameters Measurements (S11, S12, S21, S22) |
Full 2-Ports & Absolute Power Calibration | 1) Vector Network Analyzer, I/O Switchand at least one Power Meter
enabled 2) Number of VNA receivers set to 4 |
12 error terms matrix extraction from TRL, SOLT, SOLR … + Receiver Power calibration extraction from reading correct Ratioed & Unratioed Measurements | S-Parameters Measurements (S11, S12, S21, S22) 1-Tone Measurements (Pin, Pout , Gt, Gp, PAE, AMPM, Harmonics Power, …. ) 2-Tones Measurements (Pin, Pout, Gt, Gp, PAE, AMPM, C/In, OIPn, …. ) |
1-Port Enhanced Calibration | 1) Vector Network Analyzer, I/O Switch and at least one Power Meter
enabled 2) Number of VNA receivers set to 3 or 4 |
3 error terms matrix extraction from 1-Port (Reflection) calibration + Receiver Power calibration extraction from reading correct Ratioed & Unratioed Measurements | S-Parameters Measurements (only in forward S11, S21)
1-Tone Measurements (Pin, Pout, Gt, Gp, PAE, AMPM, Harmonics Power, …) 2-Tones Measurements (Pin, Pout, Gt, Gp, PAE, AMPM, C/In, OIPn, …) |