AM3100 Main Features:

  • Cost effective pulsed SMU
  • Compact and efficient design
  • Built-in power supplies
  • Pulsed/Continuous DC supply and measurement solution
  • Synchronization capabilities for multiple instruments setup
  • Drivers available for IVCAD and IQSTAR Software
  • LAN/USB remote Control using SCPI commands

Specifications

PURPOSE GATE
OPERATING RANGE
Switched voltage levels 2
Voltage ±25V
Pulsed current ±0.2A
Source DC 3W Max
Sink DC 0.5W Max
SOURCE PERFORMANCE
Voltage setting resolution 16bit 1mV
Output impedance ON: 50Ω / OFF: 50mΩ
PULSE TIMING  
Pulse Timing Pulse width from 1.1µs to 10s /20ns (Within Power limits)
PRF From 0.1Hz to 200kHz
MEASUREMENT PERFORMANCE  
V range 25V
I range 5mA/200mA
V & I Meas. Accuracy 7.5mV+0.1% ; 10µA+0.2% / 100µA+0.2%
Settling time 2-4µs
Output connector BNC

 

AM3111 Bipolar SMU +/-25V +/-200mA

The AM3111 SMU is embedded into the AM3100 controller chassis, as a bipolar pulse voltage and current generator.

PURPOSE DRAIN
OPERATING RANGE
Switched voltage levels From 0V
Voltage up to +120V
Pulsed current +30A
Average & RMS Current 5A & 10A
DC power / Pulse Power 220W / 3KW
SOURCE PERFORMANCE
Output impedance 0,1Ω
PULSE TIMING  
Pulse timing Pulse width from 1.1µs to 10s / 20ns
(Within Power limits)
PRF 0.1Hz to 100kHz
MEASUREMENT PERFORMANCE  
V range 125V
I range 33A
V & I ADC resolution 16bit
Absolute Accuracy 20mV +0.1% ; 20mA + 0.3%
Settling time 500ns to 95%, 2µs to ADC resolution
Output connector BNC

AM3121 Probe +120V +30A

The AM3121 probe is compatible with the AM3100 controller and is dedicated to bias transistor drain for general purpose measurements.

AM3111
PURPOSE GATE
OPERATING RANGE
Switched voltage levels 2
Voltage ±25V
Pulsed current ±0.2A
Source DC 3W Max
Sink DC 0.5W Max
SOURCE PERFORMANCE
Voltage setting resolution 16bit 1mV
Output impedance ON: 50Ω / OFF: 50mΩ
PULSE TIMING  
Pulse Timing Pulse width from 1.1µs to 10s /20ns (Within Power limits)
PRF From 0.1Hz to 200kHz
MEASUREMENT PERFORMANCE  
V range 25V
I range 5mA/200mA
V & I Meas. Accuracy 7.5mV+0.1% ; 10µA+0.2% / 100µA+0.2%
Settling time 2-4µs
Output connector BNC

 

AM3111 Bipolar SMU +/-25V +/-200mA

The AM3111 SMU is embedded into the AM3100 controller chassis, as a bipolar pulse voltage and current generator.

AM3121
PURPOSE DRAIN
OPERATING RANGE
Switched voltage levels From 0V
Voltage up to +120V
Pulsed current +30A
Average & RMS Current 5A & 10A
DC power / Pulse Power 220W / 3KW
SOURCE PERFORMANCE
Output impedance 0,1Ω
PULSE TIMING  
Pulse timing Pulse width from 1.1µs to 10s / 20ns
(Within Power limits)
PRF 0.1Hz to 100kHz
MEASUREMENT PERFORMANCE  
V range 125V
I range 33A
V & I ADC resolution 16bit
Absolute Accuracy 20mV +0.1% ; 20mA + 0.3%
Settling time 500ns to 95%, 2µs to ADC resolution
Output connector BNC

AM3121 Probe +120V +30A

The AM3121 probe is compatible with the AM3100 controller and is dedicated to bias transistor drain for general purpose measurements.